GENESYS™ 140 Vis Spectrophotometer
Manufacturer – Thermo Fisher Scientific
Description: GENESYS 140 is a durable, network-capable visible-range spectrophotometer built for frequent use in quality control and educational laboratories. It features a 7-inch high-resolution touchscreen and supports absorbance and transmittance measurements across 325 to 1100 nm with a 5 nm spectral bandwidth, all in a compact and reliable design.
Technical Specifications:
Wavelength range: 325 to 1100 nm
Spectral bandwidth: 5 nm
Optical design: Dual-beam
Light source: Tungsten-halogen lamp (typical lifetime >1000 hours)
Baseline flatness: ±0.005 A
Drift (at 500 nm after 1h warm-up): <0.0010 A/hr
Photometric accuracy: ±0.002 A at 0.5 A, ±0.004 A at 1.0 A, ±0.008 A at 2.0 A
Photometric repeatability: ±0.001 A at 1 A (measured at 546 nm)
Wavelength accuracy: ±0.5 nm
Wavelength repeatability: <±0.2 nm
Wavelength data interval (scan resolution): 0.2, 0.5, 1, 2, 5 nm
Noise (RMS at 500 nm, 60 consecutive measurements): ≤0.00020 A at 0 A, ≤0.00030 A at 1 A, ≤0.00040 A at 2 A
Dimensions (L × W × H): 35.5 × 38.5 × 19.5 cm
Weight: ~7.5 kg
Display: 7-inch high-definition touchscreen (800 × 1280 pixels)
Connectivity: USB-A (flash memory, duplex port), Optional Ethernet and Wi-Fi for data export
Key Features:
• Visible-only spectrophotometer covering 325–1100 nm wavelength range.
• 7-inch high-resolution touchscreen with modern, intuitive user interface.
• Dual-beam optical design ensures high stability and accurate absorbance measurements.
• USB ports (single and duplex) for data storage and peripheral devices.
• Optional Ethernet and Wi-Fi connectivity for method/data transfer and network integration.
• Compact and durable design, suitable for busy laboratory environments.
• Customizable sample compartment that accommodates a variety of accessories and cell holders.
• Supports accessories including multi-position cell changers, Peltier temperature-controlled holders, sippers, and fiber-optic probes.
• Removable, washable sample compartment liner for easy cleaning and maintenance.
• Designed for routine QA/QC, teaching, and research labs needing reliable visible-range analysis.
Applications: Used in teaching laboratories, R&D or QA/QC where visible‑region absorbance or transmission measurements are required for routine or high throughput sample analysis.
Catalogs:
840-308000
840-308300
840-308400